Series 8900 Gas Chromatograph

Click on a Product to view details.

Request Information
Series 8900 Gas Chromatograph


Series 8900 Gas Chromatograph

The Baseline® Series 8900 is a multi-purpose gas chromatograph designed to continuously monitor single or multiple gas components in a wide range of applications.

The analyzer is microprocessor based and contains powerful embedded firmware. This makes it an online instrument with GC technology that is virtually transparent to the end user. There are a variety of state of the art detector, calibration and gas train options offering the flexibility and precision necessary to analyze from sub-parts per billion (sub-ppb) to percent levels.

The 8900's automatic calibration is ideal for unattended operation. It can be set up on a bench-top or rack-mounted, and its compact size makes it transportable. There are a variety of industry ...


Series 8900GC Analyzer for Acrylonitrile in Air


Acrylonitrile in Air Application

The Series 8900 Acrylonitrile Analyzer provides direct measurement of Acrylonitrile in ambient air. This instrument is utilized in industrial plants to monitor workplace exposure limits.

The Series 8900 Acrylonitrile Analyzer employs a flame ionization detector (FID) as the sensing element. A dual column configuration with timed backflush to vent is used to strip off moisture and heavier hydrocarbons. A pre-cut column is used in series with the analytical column. At sample injection a fixed volume of sample is carried to the pre-cut column. Backflush is timed so that only the Acrylonitrile and other similar components are eluted to the analytical column. Contaminants on the pre-cut column are backflushed to vent. Acrylonitrile is ...


Series 8900GC Analyzer for BTEX in Ambient Air


BTEX in Ambient Air Application

The Series 8900 BTEX Analyzer provides direct measurement of Benzene, Toluene, Ethylbenzene, and Xylenes in ambient air. This instrument is utilized in ambient air networks around metropolitan areas and fence-line monitoring at industrial sites.

The Series 8900 BTEX Analyzer employs a photoionization detector (PID) as the sensing element. This detector is specific to volatile organic compounds. The Benzene, Toluene, Ethylbenzene, and Xylene's in the gas sample are physically separated using proprietary GC columns. A dual column configuration with timed backflush to vent is used to strip off moisture and heavier hydrocarbons. A pre-cut column is used in series with the analytical column. At sample injection a fixed volume of sample ...


Series 8900GC Analyzer for BTEX in Ambient Air using a High-Sensitivity PID


BTEX in Ambient Air Application using a High-Sensitivity PID

The Series 8900 BTEX Analyzer with a high-sensitivity PID provides direct low level measurement of Benzene, Toluene, Ethylbenzene, and Xylene's in ambient air. This instrument is utilized in ambient air networks around metropolitan areas and fence-line monitoring at industrial sites.

The Series 8900 BTEX Analyzer employs a high-sensitivity photoionization detector (PID) as the sensing element. This detector is specific to volatile organic compounds. The Benzene, Toluene, Ethylbenzene, and Xylene's in the gas sample are physically separated using proprietary GC columns. A dual column configuration with timed backflush to vent is used to strip off moisture and heavier hydrocarbons. A pre-cut column is used in series with the ...


Series 8900GC Analyzer for Diacetyl


Diacetyl Application

The Series 8900 Diacetyl Analyzer provides direct measurement of Diacetyl in ambient air. This instrument is utilized for environmental monitoring and to monitor workplace exposure limits.

The Series 8900 Analyzer for Diacetyl employs a photoionization detector (PID) as the sensing element. A dual column configuration with timed backflush to vent is used to strip off moisture and heavier hydrocarbons. A pre-cut column is used in series with the analytical column. At sample injection a fixed volume of sample is carried to the pre-cut column. Backflush is timed so that only the Diacetyl and other similar components are eluted to the analytical column. Contaminants on the pre-cut column are backflushed to vent. The Diacetyl is ...


Series 8900GC Analyzer for Formaldehyde in Air


Formaldehyde in Air Application

The Series 8900 Formaldehyde Analyzer provides direct measurement of Formaldehyde in ambient air. This instrument is utilized in industrial plants to monitor workplace exposure limits.

The Series 8900 Formaldehyde Analyzer employs a flame ionization detector (FID) as the sensing element. A dual column configuration with timed backflush to vent is used to strip off moisture and heavier hydrocarbons. A pre-cut column is used in series with the analytical column. At sample injection a fixed volume of sample is carried to the pre-cut column. Backflush is timed so that only the Formaldehyde and other similar components are eluted to the analytical column. Contaminants on the pre-cut column are backflushed to vent. Formaldehyde is ...


Series 8900GC Analyzer for Ethylene Oxide


Ethylene Oxide Application

The Series 8900 Ethylene Oxide Analyzer combines the selectivity of gas chromatography with the sensitivity of a Photoionization Detector (PID) providing interference free response to Ethylene Oxide (EtO) at levels below 0.05 ppm. This analytical method is referenced in OSHA's "Occupational Exposure to Ethylene Oxide: Final Standard" 29 CFR Part 1910.

The system documents exposure levels including: time-weighted averages (TWA), short-term exposure limits (STEL) and instantaneous alarms.

Areas that are typically monitored include; EtO Sources - gas storage areas, sterilizer doors, aeration doors, and the General Work Area - decontamination, packaging and preparation and sterile product storage areas.

The 8900 EtO ...


Series 8900GC Analyzer for Hydrogen Sulfide & Carbon Disulfide in Air


Hydrogen Sulfide & Carbon Disulfide in Air Application

The Series 8900 Hydrogen Sulfide and Carbon Disulfide Analyzer provides direct measurement of H2S and CS2 in ambient air. This instrument is utilized for continuous emission monitoring.

The Series 8900 H2S and CS2 Analyzer employs a Photoionization detector (PID) as the sensing element. A dual column configuration with timed backflush to vent is used to strip off moisture and heavier hydrocarbons. A pre-cut column is used in series with the analytical column. At sample injection a fixed volume of sample is carried to the pre-cut column. Backflush is timed so that only the H2S, CS2, and other similar components are eluted to the analytical column. Contaminants on the ...


Series 8900GC Analyzer for Methane/Non-Methane


Methane/Non-Methane Application

The Model 8900 Methane/Non-Methane Analyzer provides direct measurement of Methane and Non-Methane Hydrocarbons. This instrument is utilized in ambient air monitoring networks around metropolitan areas, fence-line monitoring at industrial plants and hazardous waste sites, and in the production of high purity industrial gases.

Methane is a naturally occurring gas generated from the decomposition of biological materials and industrial activity. Since Methane is not considered a pollutant it is the Non-Methane hydrocarbons that are of primary concern.

The Model 8900 Methane/Non-Methane Analyzer employs a flame ionization detector (FID) as the sensing element. This detector is specific to hydrocarbons. Using a single detector ...


Series 8900GC Analyzer for Methane, Methanol, & Ethanol


Methane, Methanol, & Ethanol Application

The Series 8900 provides direct measurement of methane, methanol, and ethanol for the biofuel industry.

The Series 8900 GC employs a flame ionization detector (FID). The methane, methanol, and ethanol in the gas sample are physically separated using proprietary GC columns. A dual-column configuration with timed backflush to vent is used to strip off moisture and heavier gases. At sample injection, a fixed volume of sample is carried through the pre-cut column. The backflush is timed so that primarily the MME and other similar compounds continue on to the analytical column. Contaminants are then backflushed to vent. Methane, methanol, and ethanol are separated from potentially interfering components on the analytical column and elute ...


Series 8900GC Analyzer for Methyl Bromide in Air


Methyl Bromide in Air Application

The Series 8900 Methyl Bromide Monitoring System provides an automated, direct measurement of Methyl Bromide in ambient air. This instrument can meet and exceed the requirements for ambient methyl bromide monitoring. It is utilized in fumigation facilities and industrial plants to detect potential worker exposure to methyl bromide, resulting from mechanical failure, human error or product off gassing. Multipoint sampling options allow the analyzer to monitor multiple sample locations for greater coverage.

The advantage of this system is that is both specific and sensitive to low levels of methyl bromide from less than 0.05 ppm and greater than 20ppm. It provides automatic calibration to verify accuracy. Multi level alarm setting and ...


Series 8900GC Analyzer for Methyl Ethyl Ketone, Chloroform & Carbon Tetrachloride


Methyl Ethyl Ketone, Chloroform & Carbon Tetrachloride Application

The Series 8900 Analyzer provides direct measurement of Methyl Ethyl Ketone, Chloroform and Carbon Tetrachloride in ambient air. This instrument is utilized for remediation and Super Fund sites and to monitor workplace exposure limits.

The Series 8900 Analyzer for Methyl Ethyl Ketone, Chloroform and Carbon Tetrachloride employs a flame ionization detector (FID) as the sensing element. A dual column configuration with timed backflush to vent is used to strip off moisture and heavier hydrocarbons. A pre-cut column is used in series with the analytical column. At sample injection a fixed volume of sample is carried to the pre-cut column. Backflush is timed so that only the Methyl Ethyl Ketone, Chloroform, Carbon Tetrachloride and other ...


Series 8900GC Analyzer for Naphthalene


Naphthalene Application

The Series 8900 Naphthalene Analyzer provides direct measurement of Naphthalene in ambient air. This instrument is utilized in industrial plants to monitor workplace exposure limits and at remediation and Superfund sites.

The Series 8900 Naphthalene Analyzer employs a photoionization detector (PID) as the sensing element. A dual column configuration with timed backflush to vent is used to strip off moisture and heavier hydrocarbons. A pre-cut column is used in series with the analytical column. At sample injection a fixed volume of sample is carried to the pre-cut column. Backflush is timed so that only the Naphthalene and other similar components are eluted to the analytical column. Contaminants on the pre-cut column are ...


Additional Applications


Additional Applications

Baseline - MOCON products work together to provide a total system solution for virtually any application. The Engineering Services group specializes in the manufacture of scalable, customized monitoring solutions, built to your specification. The considerable talents of this group combined with the modular design of the product line make Baseline - Mocon a leader in the development and deployment of custom solutions for in situ or portable applications.

The Baseline - MOCON Engineering Services group includes engineering, applications, and software specialists with extensive experience with gas monitoring systems.