The BevAlert® Model 9100 provides direct measurement of Acetaldehyde, Benzene, Toluene, Ethylbenzene, and Xylenes (ABTEX) in beverage-grade carbon dioxide, CO2 (ISBT Methods 11 & 12).

The instrument is utilized by specialty gas manufacturers and the food and beverage Industry to monitor trace volatile organic compounds in CO2 used in carbonated beverages. The Series 9100 BevAlert® Gas Chromatograph adheres to the stringent guidelines and methodologies established by the International Society of Beverage Technologists (ISBT) to monitor the most critical trace impurities.

The BevAlert Model 9100 employs a photoionization detector (PID). The ABTEX in the gas sample are physically separated using proprietary GC columns. A pre-cut column and an analytical column with timed backflush to vent are used to strip off moisture and heavier hydrocarbons. At sample injection, a fixed volume of sample is carried through the pre-cut column. The backflush is timed so that only the ABTEX and other similar compounds continue on to the analytical column. Contaminants on the pre-cut column are simultaneously backflushed to vent. Acetaldehyde, Benzene, Toluene, Ethylbenzene, and Xylenes are separated from potentially interfering components on the analytical column and elute to the detector for analysis.
Unit of Measure

Applications

N/A Measurement of ABTEX in beverage grade carbon dioxide.

Features

N/A

  • Direct measurement of Acetaldehyde, Benzene, Toluene, Ethylbenzene and Xylenes in CO2
  • Automatic calibration for unattended operation
  • Automatic baseline adjustment for long-term stability
  • Analog outputs for Acetaldehyde, Benzene, Toluene, Ethylbenzene, Total Xylene concentrations
  • Analog output ranges are user selectable
  • Concentration and diagnostic alarm relays
  • Graphic display of current or historical concentrations
  • Multipoint sampling options
  • RS-232 and optional LAN

  • Specifications

    Analysis Time

    N/A 720 seconds

    Detector

    N/A PID (FID or High-sensitivity PID Optional)

    Column

    N/A Capillary

    Oven Temperature

    N/A 75–85 °C, isothermal

    Carrier Gas

    N/A Hydrogen (H2), Helium (He), or Nitrogen (N2)

    Lower Detection Limit

    N/A <50 ppb Acetaldehyde, <2 ppb Benzene, <5 ppb Toluene, <5 ppb Ethylbenzene, <5 ppb Xylenes

    Accuracy

    N/A 1% of Full-scale

    Precision

    N/A 2% of Measured Value

    Span Drift (24HR)

    N/A < 2% of Full-scale

    Sample Flow Rate

    N/A 250 - 500 cc/min, typical

    Output

    N/A
    Digital: RS-232, optional Local Area Network
    Analog: (5) 0-20 mA or 4-20 mA loop power supplied, isolated. Selectable for: gas concentration, unintegrated detector signal. Options for up to 16 additional programmable 0-20 mA, 4-20 mA or voltage. outputs: 0-1 V, 0-5 V, or 0-10 V.

    Relays

    N/A (5) User programmable relays for concentration and diagnostic alarms (1A @ 30Vdc). Options for up to 32 additional relays available.

    Inputs

    N/A Digital input board for 3 contact closure inputs. Supports start analysis, start calibration, and analyze calibration gas sample.