The PetroAlert Model 8900 combines the selectivity of gas chromatography with the sensitivity, broad dynamic range, and hydrocarbon selectivity of a flame ionization detector (FID) providing fast analysis of C1-C5 hydrocarbons and Carbon Dioxide required in Mud Logging for Oil & Gas exploration.
The PetroAlert Model 8900 uses a dual column configuration with timed backflush to vent which is used to strip off moisture and C6+ hydrocarbons. A pre-cut column is used in series with the analytical column. At sample injection a fixed sample volume is carried to the pre-cut column. Backflush of the pre-cut column is timed from sample injection so that only C1-C5 hydrocarbons are eluted from the pre-cut column to the analytical column. C6+ hydrocarbons on the pre-cut column are then vented. Then a third column is used in conjunction with a second 10-port valve for individual analysis of methane. C1-C5 hydrocarbons are separated on the analytical column and elute to the flame ionization detector (FID) for analysis.
The PetroAlert Model 8900 uses a dual column configuration with timed backflush to vent which is used to strip off moisture and C6+ hydrocarbons. A pre-cut column is used in series with the analytical column. At sample injection a fixed sample volume is carried to the pre-cut column. Backflush of the pre-cut column is timed from sample injection so that only C1-C5 hydrocarbons are eluted from the pre-cut column to the analytical column. C6+ hydrocarbons on the pre-cut column are then vented. Then a third column is used in conjunction with a second 10-port valve for individual analysis of methane. C1-C5 hydrocarbons are separated on the analytical column and elute to the flame ionization detector (FID) for analysis.
Unit of Measure
Features
N/A
|
Specifications
Analysis Time |
N/A <300 seconds through Propane, <800 seconds through Pentane |
Detector |
N/A TCD |
Column |
N/A Packed |
Oven Temperature |
N/A 80ºC, Nominal |
Carrier Gas |
N/A Nitrogen, 30 cc/min, Nominal |
Lower Detection Limit |
N/A <50ppm |
Accuracy |
N/A ± 5% (relative) within dynamic range |
Precision |
N/A 2% of Measured Value |
Span Drift (24HR) |
N/A <2% of Full-scale |
Sample Flow Rate |
N/A 250 - 500 cc/min, typical |
Output |
N/A
Digital: RS-232, optional Local Area Network Analog: (1) 0-20ma or 4-20ma loop power supplied, isolated. Selectable for: gas concentration, unintegrated detector signal. Options for up to 20 additional programmable 0-20ma, 4-20ma or voltage outputs: 0-1V, 0-5V, or 0-10V. |
Relays |
N/A (5) User programmable relays for concentration and diagnostic alarms (1A @ 30Vdc). Options for up to 32 additional relays available. |
Inputs |
N/A Optional digital input board for 3 contact closure inputs. Supports start analysis, start calibration, and analyze calibration gas sample. |