The Model 8900 Methane/Non-Methane Analyzer provides direct measurement of Methane and Non-Methane Hydrocarbons. This instrument is utilized in ambient air monitoring networks around metropolitan areas, fence-line monitoring at industrial plants and hazardous waste sites, and in the production of high purity industrial gases.

Methane is a naturally occurring gas generated from the decomposition of biological materials and industrial activity. Since Methane is not considered a pollutant it is the Non-Methane hydrocarbons that are of primary concern.

The Model 8900 Methane/Non-Methane Analyzer employs a flame ionization detector (FID) as the sensing element. This detector is specific to hydrocarbons. Using a single detector configuration eliminates the detector balancing, maintenance, and additional fuel and combustion air of a dual detector analyzer. The Methane in the gas sample is physically separated from all other hydrocarbons using a proprietary GC column. It is then forwarded to the detector for analysis. The carrier gas flow is then reversed and the remaining hydrocarbon components recombine and are analyzed by the detector. Since the Methane and Non-Methane components are analyzed directly, a more accurate reading is obtained than by using a subtractive technique.
Unit of Measure

Applications

N/A Measurement of methane and non-methane hydrocarbons for environmental air quality.

Features

N/A

  • Direct Methane and Non-Methane measurement
  • Automatic baseline adjustment for long-term stability
  • Automatic FID ignition
  • Automatic calibration to a known standard
  • Automatic Hydrogen & Combustion Air shutoff for maximum safety
  • Electronic support gas regulation
  • Analog output ranges are user selectable
  • Multipoint sampling options
  • RS-232 and optional LAN

  • Specifications

    Analysis Time

    N/A <70 seconds

    Detector

    N/A Flame ionization detector (FID)

    Column

    N/A Packed

    Oven Temperature

    N/A 50ºC, Nominal

    Lower Detection Limit

    N/A 20ppb as Methane, Non-Methane 50ppb as Propane

    Accuracy

    N/A ± 5% (relative) within dynamic range

    Precision

    N/A 2% of Measured Value

    Carrier

    N/A Hydrogen, 24 cc/min typical

    Fuel

    N/A Hydrogen, 16 cc/min typical

    Combustion Air

    N/A Air, 200 cc/min typical

    Output

    N/A Analog: (1) 0-20ma or 4-20ma loop power supplied, isolated. Selectable for: gas concentration, unintegrated detector signal. Options for up to 20 additional programmable 0-20ma, 4-20ma or voltage outputs: 0-1V, 0-5V, or 0-10V.
    Digital: RS-232, optional Local Area Network

    Relays

    N/A (5) User programmable relays for concentration and diagnostic alarms (1A @ 30Vdc). Options for up to 32 additional relays available.

    Inputs

    N/A Optional digital input board for 3 contact closure inputs. Supports start analysis, start calibration, and analyze calibration gas sample.